GFUVE – GF1061 Accuracy 0.05% ( Standard Version)

GF1061 portable CT PT analyzer is mainly used for field or lab testing, it can finish the measurements (M) and protection (P) class CT, PT and TYP class CT. Adopt 7 inch touch TFT LCD, self-equipped mini type printer supporting field printing; supporting to use USB flash disk to dump data. The CT PT Analyzer is the most complete and easy-to-use testing system for protection and metering CTs according to IEEE and IEC standards.

Categories: , ,
  • Steady and transient state characteristic tests of various types of CT/PT
  • The use of advanced power technology, the test knee point reaches up to 30kV
  • No external other auxiliary equipment, stand-alone to complete all test items
  • CT test, easy to test, all the tests are using the same wire connection except the load test
  • It carries with thermal printer, so it can print test results on site
  • Parameters such as knee point current and voltage
  • Parameters such as 10% error curve, 5% error curve
  • With battery, working 48hours
  • The device can store 10000 groups of test data which would not be lost if the device is power off
  • The data can be displayed and analyzed after the test, or transferred to PC through USB disk and produce a Word file report
  • Portability: weight <4kg, the best light ct analyzer
Need more information?

Please contact +6221-5467618, +6221-54212399 or email [email protected]

Specification

I. Current Transformer (CT) II. Voltage Transformer (PT)
1. Magnetization curve 1. Excitation characteristic test
2. Transformation ratio test 2. Transformation ratio test
3. Polarity 3. Polarity
4. 5% and 10% error curve 4. Ratio error, phase error
5. Accuracy limiting factor (ALF) 5. Degauss
6. Degauss 6. Calculation of knee point value
7. Ratio error, phase error 7. Actual secondary load test (burden test)
8. Automatic calculation of excitation knee
point value
8. Resistance test
9. Actual secondary load test (burden test)
10. Resistance test
11. Secondary time constant (Ts)
12. Remanence coefficient (Kr)
13. Transient dimensioning factor (Ktd)
14. Peak instantaneous error (Er)
15. Magnetizing inductance (LU)
16. Instruments security factor(FS)
17. Composite error

Downloads