HIOKI – 3535

  • Wide range from 100kHz to 120MHz
  • High speed LCR testing (6ms/sample)

High-speed LCR meter with up to 120MHZ sampling

  • Wide range from 100kHz to 120MHz
  • High speed LCR testing (6ms/sample)
  • Removable head amplifier
  • “Load compensation function” for comparing standard component and providing compensation
Need more information?

Please contact +6221-5467618, +6221-54212399 or email [email protected]

Specification

Measurement Items |Z|, |Y|, Q, Rp, Rs(ESR), G, X, B, Θ, Ls, Lp, Cs, Cp, D(tan δ)
Measurement ranges

1kΩ range 10kΩ range 100kΩ range
Z, R 100mΩ to 2kΩ 1kΩ to 20kΩ 10kΩ to300kΩ
C 0.66pF to 15.9µF 0.066pF to 1.59nF 4.4fF to 159pF
L 0.133nH to 3.18mH 1.33µH to 31.8mH 13.3µH to 477mH
Θ -180.00° to 180.00°

Measurement Frequency 100 kHz to 120 MHz ( 100Hz to 100kHz steps )
Basic accuracy |Z|: ±0.5 % rdg. , phase angle: ±0.3°
Output Impedance 50 ±10 ohm (at 100 kHz)
Measurement Signal Level 5 mV to 1 V, 20 mA max. ( up to 10.00 MHz )
5 mV to 500 mV, 10 mA max. ( above 10.01 MHz )
Setting Resolution: 1mV steps

200 µA to 20 mA: 1 V max. ( up to 10.00 MHz )
200 µA to 10 mA: 0.5 V max. (above 10.01 MHz)
Setting Resolution: 10 µA steps

Measurement Time 6 ± 1 ms (nominal)
Other function Monitor Function, Limit Function, Average Trigger Function, Load Compensation Function, Key Lock Function, Comparator, Classification ( BIN ) Measurement, Correlation, Compensation Function, Panel Save and Load, Measurement Value Storage, Zoom Display Function, Continuous Measurements, Audible Beeper, Numerical Display, Digit Setting Function, Display Setting Functions, Printer Function
Interfaces GP-IB, RS-232C and EXT I/O ( standard )
Power 100 to 240 VAC, 50/60 Hz, 50 VA
Dimensions and Mass Approx. 360W×130H×360D mm, 8.3 kg
Supplied accessories Power cord