HIOKI – 8205-10

The 8205-10 can be used for basic recording in a wide range of applications, with features such as a wide AC/DC measurement range of 0.1V to 500V, an analog level monitor function, and a special clamp-on probe for recording large currents.

 

The MICRO HiCORDERs 8205-10 and 8206-10 incorporate the latest technology in HIOKI’s best selling MICRO HiCORDER 8200 series. Maintaining the operational simplicity of an analog multimeter, these data recorders offer easy to use features such as :

  • A fast LCD level meter vs. a stylus to quickly display the measured level
  • True RMS rectification system to accurately read distorted waveforms
  • Climate-resistant recording paper is easy to read and lasts longer!
  • Sampling rate of 100 samples/second with selectable chart speeds

 

Need more information?

Please contact +6221-5467618, +6221-54212399 or email [email protected]

Specification

Analog input UNIVERSAL UNIT LR8501
Actual functionality will depend on combination of units installed
[No. of channels] 15 analog channels; isolated scanning method input (4 terminals: push-button type)
[Voltage] 10 mV to 100 V, 1-5 V f.s. 10 ranges, Max. resolution: 500 nV
[Temperature: Thermocouples] -200°C to 2000°C (depends on sensor), 3 ranges (K, J, E, T, N, R, S, B, W), Max. resolution 0.01°C
[Temperature: Pt 100, JPt 100 sensor] -200°C to 800°C, 3 ranges, Max. resolution 0.01°C
[Resistance] 10 Ω to 200 Ω f.s. 4 ranges, Max. resolution 0.5 mΩ
[Humidity] 100 % rh f.s. 5.0 to 95.0 % rh (when using Z2000), resolution 0.1 % rh
[Max. rated voltage between isolated input channels] 300 V DC
[Max. allowable input] ±100 V DC
[Max. rated voltage from isolated terminals to ground] 300 V AC, DC
Caution: Not isolated from each other and common GND at platinum temperature sensor or resistance measurement input
Analog input VOLTAGE/TEMP UNIT LR8500
Actual functionality will depend on combination of units installed
[No. of channels] 15 analog channels; isolated scanning method input (2 terminals: M3 screw type)
[Voltage] 10 mV to 100 V, 1-5 V f.s. 10 ranges, Max. resolution: 500 nV
[Temperature: Thermocouples] -200°C to 2000°C (depends on sensor), 3 ranges (K, J, E, T, N, R, S, B, W), Max. resolution 0.01°C
[Temperature: Pt 100, JPt 100 sensor] [Resistance] not available
[Humidity] 100 % rh f.s. 5.0 to 95.0 % rh (when using Z2000), resolution 0.1 % rh
[Max. rated voltage between isolated input channels] 250 V DC
[Max. allowable input] ±100 V DC
[Max. rated voltage from isolated terminals to ground] 300 V AC, DC
Caution: Max. voltage from terminals to ground without damage
Pulse, Digital input [No. of channels] 8 channels, pulse / digital selectable for each channel, M3 screw terminal, not isolated, common ground
[Pulse totalization] 0 to 1000 M pulse, 1 range (No-voltage ‘a’ contact; normally open, open collector or voltage input), Max. resolution 1 pulse
[Pulse revolutions] 0 to 5000 /n (r/s) f.s. 1 range (same as Pulse totalization input signal condition), resolution 1/n (r/s) Note: “n” is the number of sensor output pulses per revolution, 1 to 1000
[Digital input] Record logical “1” or “0” at each sampling
[Max. rated voltage between input channels] Not isolated
[Max. allowable input] 0 to 50 V
Recording intervals 10 ms to 50 ms, 100 ms to 1 hour, 19 selections (All input channels are scanned at high speed during every recording interval)
Note: limited by using channels at 10 ms to 50 ms interval
Digital filter Select from OFF/ 50 Hz/ 60 Hz (the cut-off frequency is automatically set)
Data storage Internal memory: 8 M-words, Data storage media: CF card or USB memory
(Only data recorded to a genuine HIOKI CF card is guaranteed)
LAN interface 100BASE-TX, Functions: Data acquisition using bundled software or PC commands, FTP server, FTP client, HTTP server function, or E-mail system
Note: LAN communication support planned from software Ver. 1.20
USB interface USB 2.0 High-speed capable, series mini-B receptacle
Functions: Data acquisition using bundled software or PC commands, Transfer data from the CF card to a PC via USB drive mode
Note: Data transfer not possible from USB memory sticks
Display device 5.7 inch TFT color liquid crystal display (640 × 480 pixel)
Other functions Save waveform data in real time to the CF card or USB memory stick, Numerical value calculations, Waveform calculations, and others
AC Power Using the AC ADAPTER 9418-15 (100 to 240 V AC, 50/60 Hz), 7 VA
DC Power Using the BATTERY PACK Z1000 (optional accessory), Continuous use 5 hr, External power: 10 to 28 V DC (Please contact your HIOKI distributor for connection cord)
Dimensions & Mass 272 mm (10.71 in) W × 182.4 mm (7.18 in) H × 66.5 mm (2.62 in) D, 1.8 kg (63.5 oz), (LR8400-20 main unit, excluding the Battery Pack 370 g/ 13.1 oz)
Accessories Detailed operating manual ×1, Measurement guide ×1, AC ADAPTER 9418-15 ×1, USB cable ×1, CD-R (data collection software “Logger Utility”) ×1