TIME ELECTRONICS – 5077 Power Calibrator

Supplied with its own “PC Virtual Control” software allowing the user full control via laptop or PC. Frequency can be set in 0.1Hz steps from 45 to 400Hz, and phase in 0.1degree steps.

Supplied with its own “PC Virtual Control” software allowing the user full control via laptop or PC. Frequency can be set in 0.1Hz steps from 45 to 400Hz, and phase in 0.1degree steps. Alternatively power factor (PF) can be set in 0.01 steps. The output can be displayed as VA or Watts.

The standard maximum AC current output (22A) can be extended to 100A by using an optional 100A current transformer. This extends the maximum power to 0.1MW/MVA. A clamp meter adaptor option is also available (1 and 50 turns).

EasyCal – Manage, Automate and Optimise the Calibration Process
The 5077 can also be controlled using Time Electronics’ EasyCal software. This provides increased speed of calibration and consistency of results. Easily produce calibration certificates and reports to ISO 9001,ISO 17025, and other international quality standards.

Need more information?

Please contact +6221-5467618, +6221-54212399 or email [email protected]

Specification

Technical Specifications

DC Voltage
Range Accuracy 1 year Output Resistance Max Output Current Resolution
*2V 0.01% + 500μV < 0.5Ω 20mA 1μV
*20V 0.01% + 500μV < 0.5Ω 20mA 10μV
300V 0.02% + 30mV < 5Ω 20mA 100μV
1050V 0.05% + 50mV < 10Ω 10mA 1mV

*Accuracy is specified at minimum current output.

DC Current
Range Accuracy 1 year Compliance Voltage Resolution
2A 0.03% + 500μA 5V 100μA
22A 0.05% + 6mA 4V 1mA

 

AC Voltage (45Hz to 400Hz 0.02% accuracy)
Range Accuracy 1 year Output Resistance Max Output Current Resolution
*2V 0.03% + 500μV < 0.5Ω 20mA 100μV
*20V 0.03% + 2mV < 1Ω 20mA 1mV
**300V 0.06% + 30mV < 5Ω 20mA 10mV
**1050V 0.08% + 90mV < 10Ω 10mA 100mV

*Accuracy is specified at minimum current output.
**Accuracy is specified at 100Hz for current outputs less than 5A.

AC Current (45Hz to 400Hz 0.02% Accuracy)
Range Accuracy 1 year Compliance Voltage Resolution
2A 0.1% + 2mA 3.5V 100μA
22A 0.1% + 20mA 3V 1mA

 

Phase Angle (Applies to outputs above 22 Volts)
Frequency Range Accuracy 1 year Resolution
45 to 99Hz -90.0 to +90.0 deg 0.3 deg 0.1 deg
100Hz to 400Hz -90.0 to +90.0 deg 1 deg 0.1 deg

 

Power Factor
Frequency Range Resolution
45 to 400Hz 0.00 to 1.00 0.01

The accuracy of the power is complex and is determined by using a formula, which combines the errors due to Voltage, Current, and Phase.
Power Accuracy (%) = SqrRt (Voltage accuracy2 + Current accuracy2 + Phase Correction2)
Where Phase Correction (%) = 100x(1-Cos(Phase+PhaseAcc)/Cos Phase)